Hyperbaric/Hypobaric Chamber

  • Hypobaric Chamber ProOx-830
Hypobaric Chamber ProOx-830

Hypobaric Chamber ProOx-830

  • Product description: Hypobaric Chamber ProOx-830 for animal experiment is used to simulate the hypobaric plateau environment. The pressure can be set according to research requirements. The maximum altitude can be simulat
  • INQUIRY

Hypobaric Chamber ProOx-830

        Hypobaric Chamber ProOx-830 for animal experiment is used to simulate the hypobaric plateau environment. The pressure can be set according to research requirements. The maximum altitude can be simulated 10,000 meters above sea level. It can also be used for constant hypoxia control. It is suitable for small to medium-size animal like dogs, monkeys, rabbits, murine, etc.

Product Features and Parameters

Safety Measure

From the beginning of design to the final completion of testing and inspection, all experimental oxygen chambers comply with strict manufacturing practice.

Equipped with automatic ventilation function, and comply with Animal Husbandry Code.

Animal Chamber

The chamber is made of imported transparent PMMA, which is strong and durable for facilitate observation.

Supported by aluminum alloy and stainless-steel material.

Casters with brake design for safety and portability.

Multifunctional Control

Constant Hypobaric Hypoxia Mode

Constant Normobaric Hypoxia Mode

Dynamic Intermittent Hypoxia Mode

Dynamic Multistage Control Mode

Experimental Monitoring, Control, Alarm

Monitoring parameters: chamber temperature, humidity, oxygen, pressure, simulated altitude, carbon dioxide (optional);

Professional imported oxygen sensor, monitoring range: 0.1-25.0%;

Monitoring and control function, 10-inch touch screen control, real-time display of pressure and oxygen concentration dynamic change curve, and monitor of the experimental process;

2 control modes are available: constant oxygen concentration mode, and intermittent oxygen concentration mod. Various parameters can be set, for example, upper limit oxygen concentration, lower limit oxygen concentration, upper limit oxygen concentration maintenance time, lower limit oxygen concentration maintenance time, oxygen filling time, nitrogen filling time, and etc.

Control parameters: target pressure, simulated altitude rising time, hypobaric maintenance time, simulated altitude declination time, atmospheric pressure maintenance time, gas flow rate, ventilation rate, duration of each ventilation, ventilation gas flow rate, oxygen concentration, etc.

Alarm parameters: temperature upper limit, temperature lower limit, pressure upper limit, pressure lower limit, oxygen concentration upper limit, oxygen concentration lower limit.

Our company can provide 3Q qualification to assist customers to qualify the system in all aspects.

Scan the qr codeClose